Select Store:
Select Currency:

Sick LFP1200-G1NMB (1052077)

Product Summary

HYGIENIC LEVEL PROBE
Manufacturer: Sick
Manufacturer part number: 1052077
VS ID: 185376
Weight(kg): 2.5000

Click here to check availability

Pricing - Condition Manufacturer New

$920.95 excl tax
$874.90 excl tax
  • Level monitoring in hygienic applications
  • Manually cutable monoprobe made from stainless steel 1.4404 up to 2,000 mm long with Ra ≤ 0.8 µm
  • Process temperature up to 150 °C, process pressure up to 16 bar
  • CIP/SIP resistant
  • IP 67 and IP 69K enclosure rating
  • Interchangeable hygienic process connections
  • 3-in-1: combined display, analog output and binary output
  • Analog output 4 mA ... 20 mA / 0 V … 10 V, switchable, plus two binary outputs
  • Your benefits

  • Robust design increases service life
  • High flexibility due to cutable probe and interchangeable connection concept
  • Cost savings due to multiple output signals: one system for both level detection and continuous level monitoring
  • Time and cost savings due to low maintenance and quick commissioning
  • No calibration or recalibration required for commissioning, thus saving time and costs

Features
Medium:Fluids
Measurement:Switch, Continuous
Probe length:1,200 mm
Process pressure:-1 bar ... 16 bar
Process temperature:-20 °C ... 150 °C
EHEDG approval:
GOST approval:
RoHS certificate:
IO-Link:
3-A:0
Performance
Resolution:< 2 mm
Accuracy of sensor element:± 5 mm
Repeatability:≤ 2 mm
Dielectricity constant:≥ 5
Inactive area at process connector:25 mm 1)
Inactive area at probe end:10 mm 2)
Conductivity:No limitation
Response time:< 400 ms
1) With parameterized tank with water under reference conditions, otherwise 40 mm  2) With water under reference conditions  
Mechanical data
Housing material:303
Max.probe load:6 Nm
Process connection:G ¾ A
Wetted parts:316L (Ra ≤ 0,8 µm), PEEK
Housing design:With PMMA viewing window
Electrical data
Temperature drift:< 0.1 mm/K
Enclosure rating:IP 67: EN 60529, IP 69K: EN 40050
Supply voltage:12 V DC ... 30 V DC 1)
Power consumption:≤ 75 mA at 24 V DC without output load
Protection class:III
Initialization time:≤ 2 s
Output load:0 V ... 10 V > 750 Ohm at Uv >= 14 V, 4 mA ... 20 mA < 400 Ohm at Uv > 12 V, 4 mA ... 20 mA < 500 Ohm at Uv > 13,5 V
Upper signal level:20 mA ... 20.5 mA
Lower signal level:3.8 mA ... 4 mA
Hysteresis:Min. 2 mm, free adjustable
Output current:< 100 mA
Signal voltage HIGH:Vs - 2 V
Signal voltage LOW:≤ 2 V
Capacitive load:100 nF
Inductive load:< 1 H
Connection type:Round connector M12 x 1, 5-pin
EMC:EN 61326-1:2006, 2004/108/EG
Interference resistance:EN 61000-6-2:2005
Interference emission:EN 61000-6-4:2007
Single- and continuous shock:EN 60068-2-27
1) All connections are polarity protected. All outputs are overload and short-circuit protected.  
Ambient data
Ambient operating temperature:-20 °C ... 60 °C
Ambient storage temperature:-40 °C ... 80 °C
Products specifications
Manufacturer category Sick LFP
Manufacturer Sick